Test methods and records on machined medical parts

A machined implant or instrument is judged on tests run on the bar, tests run on the part and the records that tie both to one lot. This index lists the test methods that appear on drawings and certificates, which standard sets each one, what the report must show and who usually runs it.

Records this lot must leave behind

Every test on this page ends in a record. The lot file for a machined medical part holds these.

RecordWhat it showsUsual source
Mill certificateSpecification, heat number, chemistry, tensile results and any metallurgical resultsMill, copied by the distributor
Laboratory reportOne test, one method and edition, specimen, result, signatoryMill or accredited laboratory
Process recordsHeat treatment, passivation, cleaning and marking, each with its own validationShop or sub-tier vendor
Inspection recordsDimensional and surface results against the drawingShop
TravelerThe route, with the heat number carried through every operationShop

The test methods

The table lists the methods most often cited on implant and instrument drawings. Standards are named for orientation. Check the edition on the drawing, and read the pass limit in the material specification, since the test standard sets none.

TestWhat it showsMethod documentsWhere to read more
ChemistryThe alloy is within its limits, element by elementRecognized analytical procedures named by the lab. For titanium, ASTM E1409 (oxygen and nitrogen), E1447 (hydrogen) and E1941 (carbon) are the usual inert gas fusion and combustion methodsChemistry reports for Ti-6Al-4V ELI
Tensile and yieldStrength, yield strength, elongationISO 6892-1, which ISO 5832 cites; ASTM tension methods on the ASTM sideASTM F1472
HardnessCondition of the material; Brinell is named in the ASTM F138 abstractBrinell, Rockwell, Vickers; microindentation (ASTM E384) for near-surface workHardness and microhardness
Grain sizeAverage grain size number of a single-phase structureASTM E112 or ISO 643Grain size and inclusion rating
Inclusion ratingCleanliness of wrought steelASTM E45 or ISO 4967Grain size and inclusion rating
Microstructure of titanium barFine alpha and beta mix, no continuous alpha network at prior beta boundariesText of the material specification; ISO 20160 photomicrographs for bar under ISO 5832-3ASTM F136
Alpha caseOxygen-rich surface layer from heat in airPolished and etched section, hardness traverse; ASTM E3 and E407 for preparation and etching, E384 for hardnessAlpha case
Corrosion of instrumentsResistance of stainless instruments to boiling, copper sulfate, autoclaving and thermal exposureASTM F1089, ISO 13402ISO 13402 and F1089
Corrosion of small implantsSusceptibility of a finished small implant in test solutionASTM F2129, cyclic potentiodynamic polarization; ASTM F746 for pitting screeningCorrosion testing
Passivation and cleanlinessFree iron removed, residues within limitsASTM A967, ASTM F86, ISO 19227 for orthopedic implantsASTM F86
Cleaning validationResidues of machining brought below the device maker's limit by a fixed processASTM F3127, ASTM F2459, ISO 19227Cleaning validation steps
Bioburden and endotoxinMicrobial load and endotoxin release of the device, run by the device maker's laboratoryISO 11737-1, ISO 11737-3, USP 85 and 161, FDA Edition 2 guidanceBioburden and endotoxin
Metallography laboratorySections, etch and structure read by someone competent, in the shop or outsideASTM E3, E407, E112, E45, E384; ISO 20160 for titanium barIn-house or outsourced
Matrix of implant materials against test methods: chemistry, tensile, microstructure, grain size, inclusions, alpha case and corrosion
Which checks apply to which material. Typical reading, confirm with the specification on the drawing.

Which checks apply to which material

Material and documentStructure ruleCleanliness and grainOther checks
Ti-6Al-4V ELI, ASTM F136Fine alpha and beta, no continuous alpha networkNot graded by grain size numberAlpha case where the surface has seen heat
Ti-6Al-4V, ASTM F1472Same wording as F136Not graded by grain size numberAlpha case, ISO 20160 for bar
316LVM, ASTM F138No delta ferrite, chi or sigma phaseGrain size and inclusion rating, limits in the edition and in ISO 5832-1Passivation after machining
CoCrMo, ASTM F1537Homogeneous structureGrain size ASTM No. 5 or finer, as reportedCondition and tensile results
MP35N, ASTM F562Check the editionCheck the editionCondition: annealed, cold worked or aged
Instrument stainless, ASTM F899Per the gradePer the gradeBoil test, copper sulfate test, autoclave resistance

Who runs which test

The device maker chooses the tests and writes the limits. The mill and its laboratory run the material tests and report them on the certificate. The shop runs its own dimensional inspection and the process steps that need records, and keeps the rest with the lot. The shop does not decide equivalents, accept out-of-limit results or re-test bar on its own authority. Where a contract does ask for a test at receiving, the quality agreement should name it.

How to read any test report

  1. Match the heat or lot number to the certificate and the traveler.
  2. Find the method and its edition. A result without a method cannot be compared.
  3. Check the specimen. Where it was taken, in what condition, in which direction.
  4. Find the limit and its source. The document and clause the result was judged against.
  5. Check who signed and who accredited. A named signatory, and the method on the laboratory scope where the device file asks for it.

How the detail pages relate

Each detail page takes one test and sets out what the standard measures, which specifications call for it, what a report must show and what the shop is asked to do. Material pages in the material specifications index give the limits. Regulation and quality-system pages are in the standards index. The inspection plan guide covers dimensional checks and sampling.

Information, not regulatory advice. Qualification stays with the device maker, inside its own quality system. Editions change: check the current edition of each document at its source.

Frequently asked questions

Who runs these tests, the shop or the mill?

Most material tests, such as chemistry, tensile strength, grain size and inclusion rating, are run by the mill or by an accredited laboratory and reported on the certificate. A machining shop receives the report, checks it against the drawing and files it with the lot. The shop runs the tests that belong to its own processes, such as dimensional inspection, and may run others if the contract says so.

Does a test report prove the finished part is good?

It proves what it tested. A report on the bar describes the bar, and ISO 5832 itself warns that a sample taken from a finished product can differ from the tabulated values. Finished-part evidence comes from the inspection, cleaning and process validation records.

What should a test report always show?

The heat or lot number, the method and its edition, the specimen and where it was taken, the result, the limit it was judged against and a named signatory. A laboratory report should also show that the method is on the laboratory's accreditation scope when the device file requires accredited testing.

Where do the acceptance limits come from?

From the material specification, the drawing or the device maker's purchase specification. The test standards themselves, such as ASTM E112 and E45, describe how to measure and set no limits.

Sources

Standards and regulations change. Check the current text at the source before you rely on it in a submission.

Last reviewed 19 October 2026