Guides / Materials
Titanium, 316L and PEEK: Choosing a Material for a Machined Medical Part
September 23, 2026
Material choice on a machined medical part is usually made by the regulatory file rather than by the machinist, and that is correct. But the way it is written on the drawing decides what you pay, and a surprising number of drawings leave money on the table by being imprecise in one specific way.
Start from the requirement, not the alloy
Five questions rule material in and out before anyone thinks about machining.
- Patient contact, and for how long? Transient, prolonged or permanent contact puts you in very different biocompatibility territory.
- Load, and is it cyclic? A cyclically loaded implant cares about fatigue and fracture toughness in a way a static part does not.
- Imaging. Metal artefact in CT or MRI can be the reason a polymer wins.
- Sterilisation method. Autoclave cycles are brutal on some stainless grades and on some polymers.
- Service life. A single use instrument and a twenty year implant are different problems.
Answer those and the material list is usually two items long. Then the question becomes what to write down.
Titanium
Ti-6Al-4V ELI, Grade 23, to ASTM F136. The workhorse for load bearing implants: trauma plates, bone screws, spinal rods, stems. ELI means extra low interstitials, and in practice it means tighter limits on oxygen and iron than standard Grade 5. That buys better fracture toughness and better fatigue crack growth resistance, which is the whole point on a part that will be cycled for years.
Here is the part worth knowing. Grade 23 and Grade 5 machine essentially identically. Same speeds, same feeds, same tooling, same coolant. The premium on ELI is the bar and the certificate, not the cutting.
So when a quote comes back with a large machining uplift for ELI, that is a fair question to ask. The material line should move. The hours line should not move much.
CP titanium, Grade 2 and Grade 4, to ASTM F67. Commercially pure. Grade 2 is soft and formable and shows up in dental abutments, housings and cranial mesh. It is gummier than Grade 23 on the machine: it smears and builds up on the cutting edge rather than breaking the chip, so finish and burr control take the time rather than the roughing.
Titanium of any grade has low thermal conductivity, which means the heat that would leave in the chip stays in the tool instead. That is why surface speeds look slow to anyone used to aluminium, and why tooling is a real line in a titanium quote.
Stainless
316L and 316LVM to ASTM F138. The LVM is vacuum remelted and the F138 designation is the implant bar with a certificate behind it. General 316L from a stockist is a different purchase. It work hardens fast, which has a practical consequence: a tool that rubs once makes the next pass harder, so a shop that is guessing at feeds will fight the material and pass the cost to you. It needs passivation after machining and it shows every scratch on a polished face.
17-4 PH, condition H900 or H1150. Reusable instrument bodies, jaws, drivers. Not an implant grade. The heat treat condition changes the part substantially and must be on the drawing. “17-4 PH” on its own is an incomplete specification and different shops will make you different parts in good faith.
Martensitic grades for cutting edges that have to survive several hundred autoclave cycles without pitting.
Cobalt chrome
CoCrMo to ASTM F75 cast or F1537 wrought, and ISO 5832-12. Knee femoral components, hip heads, dental frameworks. It wins on wear where titanium does not.
On the machine it is the hardest of the group on a tool, and the cost driver is tool life rather than machine time. Much of the finishing is ground or polished rather than cut. If you have a CoCr part with a mirror bearing surface, most of your quote is the surface, not the geometry.
Polymers
Implant grade PEEK to ASTM F2026. Spinal interbody cages, trials, instrument handles. It cuts easily, which is exactly how it catches people out. The risks are heat and clamping stress causing crazing, and contamination from a machine that has been running steel. A shop that runs implant PEEK properly treats it as a segregated job with its own tooling, and that is what you are paying for rather than the cycle time.
UHMWPE to ASTM F648 and ISO 5834-2. Knee inserts and acetabular liners. Very low melting point, springs back off the tool, needs sharp tooling and light cuts. Measuring it accurately is its own discipline because the part moves while you gauge it.
Nitinol
ASTM F2063. Guide wires, stents, staples, flexible instruments. Superelastic, so it moves away from the tool and springs back. The honest answer is often that it is ground, laser cut or EDM cut rather than milled. Ask before you assume a mill will do it, because a shop that has not run it before will find out on your part.
The comparison
| Material | Standard | Where it goes | The cost driver in machining |
|---|---|---|---|
| Ti-6Al-4V ELI, Gr 23 | ASTM F136, ISO 5832-3 | Load bearing implants, trauma | Tooling, heat in the cutting edge |
| CP Ti, Gr 2 and 4 | ASTM F67, ISO 5832-2 | Dental, housings, mesh | Burr control and finish, not roughing |
| 316L / 316LVM | ASTM F138, F139 | Instruments, fluidics, temporary fixation | Work hardening, then passivation |
| 17-4 PH | ASTM A564, AMS 5643 | Reusable instrument bodies | Heat treat condition, hardness at cut |
| CoCrMo | ASTM F75, F1537 | Knee, hip, dental frameworks | Tool life and the polishing route |
| PEEK, implant grade | ASTM F2026 | Spinal cages, trials, handles | Segregation, contamination control |
| UHMWPE | ASTM F648, ISO 5834-2 | Knee inserts, liners | Sharp tooling, and metrology |
| Nitinol | ASTM F2063 | Wires, stents, staples | Often not milled at all |
What to write on the drawing
Four things, and the second one is the one people skip.
- The material and its standard. Ti-6Al-4V ELI to ASTM F136, not “titanium”.
- The condition, where it applies. 17-4 PH H900 is a different part from 17-4 PH H1150.
- The surface treatment and its standard. Passivate to ASTM A967, electropolish to ASTM B912, and say which method where the standard offers several.
- The acceptance criterion. How you will judge it, so the shop knows what it is being measured against before it quotes rather than after it ships.
And one thing worth doing before you send anything out: go down the material callouts and ask which were chosen for this part and which were copied from the last drawing. On non-implant parts, inherited implant grade specifications are one of the most common places a programme pays a premium for nothing.
If you want the quote side of this, what drives the cost of a machined medical component covers the other five drivers.
Sources
- ASTM F136, wrought Ti-6Al-4V ELI alloy for surgical implant applications
- ASTM F138, wrought 18Cr-14Ni-2.5Mo stainless steel bar and wire for surgical implants
- ASTM F2026, polyetheretherketone (PEEK) polymers for surgical implant applications
Standards and regulations change. Check the current text at the source before you rely on it in a submission.
Frequently asked questions
Does Ti-6Al-4V ELI machine differently from Grade 5?
No. Speeds, feeds, tooling and coolant are effectively the same. The difference between them is chemistry and certification, not cutting behaviour. If a shop quotes a large machining premium for ELI over Grade 5, ask what the premium is actually for, because the answer should be the bar and the traceability rather than the time.
Can I substitute Grade 5 titanium for ELI to save money?
Not on a load bearing implant. ELI to ASTM F136 exists because tighter oxygen and iron limits give better fracture toughness and fatigue crack growth resistance, which is exactly what a cyclically loaded implant needs. On a non-implant part, ask whether ELI was specified for a reason or inherited from another drawing.
Why do shops charge more for PEEK than the material cost suggests?
Because contamination and stress are the risk, not the cutting. Implant grade PEEK needs clean tooling and fixturing that has not just run steel, controlled clamping to avoid crazing, and traceability back to a certified lot. The machine time is the cheap part.
Do I put the alloy or the standard on the drawing?
Both, with the standard doing the work. Ti-6Al-4V is an alloy. Ti-6Al-4V ELI to ASTM F136 is a purchase, with a certificate tying the bar to a heat. Writing only the alloy name invites a supplier to quote the cheaper material in good faith.